User manual
Remote Control
R&S
®
RTO
484User Manual 1316.0827.02 ─ 06
TRIGger<m>:HOLDoff:MAX............................................................................................487
TRIGger<m>:HOLDoff:AUTotime.....................................................................................488
TRIGger<m>:HOLDoff:SCALing......................................................................................488
TRIGger<m>:SEQuence:MODE <Type>
Selects the type of the sequence.
See also: chapter 3.3.4, "Sequence", on page 80.
Suffix:
<m>
.
1..3
The suffix is irrelevant.
Parameters:
<Type> AONLy | ABR
AONLy
Triggers only on A-events. Additionally, a holdoff condition can be
set.
ABR
Triggers if all conditions of A- and B-events, as well as additional
delay and reset timeout or R-event (reset) conditions are fulfilled.
*RST: AONLy
TRIGger<m>:SEQuence:DELay <Delay>
Sets the time the instrument waits after an A-event until it recognizes B-events.
Suffix:
<m>
.
1..3
The suffix is irrelevant.
Parameters:
<Delay> Range: 0 to 50
Increment: 1E-12
*RST: 0
Default unit: s
TRIGger<m>:SEQuence:COUNt <Events>
Sets the number of B-events to be fulfilled after an A-event. The last B-event causes the
trigger.
Suffix:
<m>
.
1..3
The suffix is irrelevant.
Parameters:
<Events> Range: 1 to 2147483647
Increment: 10
*RST: 1
Command Reference