User manual
Remote Control
R&S
®
RTO
482User Manual 1316.0827.02 ─ 06
TRIGger<m>:LEVel<n>:NOISe:MODE.............................................................................482
TRIGger<m>:LEVel<n>:NOISe:ABSolute.........................................................................483
TRIGger<m>:LEVel<n>:NOISe:RELative..........................................................................483
TRIGger<m>:LEVel<n>:NOISe[:STATe] <HysteresisMode>
Selects how the hysteresis is set.
Suffix:
<m>
.
1..3
Event in a trigger sequence, the suffix is irrelevant.
<n> 1..9
Indicates the trigger source:
1...4 = channel 1...4
5 = External Trigger Input on the rear panel for analog signals
6...9 = Not available
Parameters:
<HysteresisMode> AUTO | MANual
AUTO
This is the recommended mode. The hysteresis is set by the
instrument to reject at least the internal noise of the instrument.
You can define a higher minimum value using TRIGger<m>:
LEVel<n>:NOISe:ABSolute.
MANual
The hysteresis is defined directly with TRIGger<m>:
LEVel<n>:NOISe:ABSolute.
*RST: AUTO
TRIGger<m>:LEVel<n>:NOISe:MODE <HystSizeMode>
Selects how the hysteresis is set.
Suffix:
<m>
.
1..3
Event in a trigger sequence, the suffix is irrelevant.
<n> 1..9
Indicates the trigger source: 1...4 = channel 1...4;
5 = External Trigger Input on the rear panel;
6...9 = Not available
Parameters:
<HystSizeMode> ABS | REL
ABS
The hysteresis is set in absolute values (voltage).
REL
The hysteresis is defined in relative values (div).
*RST: ABS
Command Reference