R&S
®
RTO
481User Manual 1316.0827.02 ─ 06
●
D: CH4
Suffix:
<m>
.
1..3
Event in a trigger sequence: 1 = A-event, 2 = B-event.
<n> 1..15
Defines the trigger type, see table 16-13.
Parameters:
<Operator> DIRect | NOT
DIRect
Input value remains unchanged
NOT
Input value is inverted
*RST: DIRect
TRIGger<m>:
QUALify<n>:AB:LOGic <Operator>
TRIGger<m>:QUALify<n>:CD:LOGic <Operator>
TRIGger<m>:QUALify<n>:ABCD:LOGic <Operator>
Defines the logical combination of the indicated channels after evaluating the previous
logical operations:
●
AB: CH1 and CH2
●
CD: CH3 and CH4
●
ABCD: all four channels
Suffix:
<m>
.
1..3
Event in a trigger sequence: 1 = A-event, 2 = B-event.
<n> 1..15
Defines the trigger type, see table 16-13.
Parameters:
<Operator> AND | NAND | OR | NOR
AND
logical AND, conjunctive combination
NAND
logical NOT AND
OR
logical OR, disjunctive combination
NOR
logical NOT OR
*RST: AND
16.2.6.3 Noise Reject
TRIGger<m>:LEVel<n>:NOISe[:STATe]...........................................................................482