User manual
Remote Control
R&S
®
RTO
480User Manual 1316.0827.02 ─ 06
Parameters:
<AddTrigLogic> ON | OFF
ON
The qualification expression is considered for the trigger event.
OFF
The qualification expression is ignored for the trigger event.
*RST: OFF
TRIGger<m>:QUALify<n>:A[:ENABle] <State>
TRIGger<m>:QUALify<n>:B[:ENABle] <State>
TRIGger<m>:QUALify<n>:C[:ENABle] <State>
TRIGger<m>:QUALify<n>:D[:ENABle] <State>
Selects one of the channels and the specified trigger event to be considered for qualifi-
cation:
●
A[:ENABle]: CH1
●
B[:ENABle]: CH2
●
C[:ENABle]: CH3
●
D[:ENABle]: CH4
You can select all channel signals except for the trigger source. In pattern trigger setup,
the trigger source channel is selected by default, and you can select all other channel
signals.
Suffix:
<m>
.
1..3
Event in a trigger sequence: 1 = A-event, 2 = B-event.
<n> 1..15
Defines the trigger type, see table 16-13.
Parameters:
<State> ON | OFF
ON
The qualification expression is considered.
OFF
The qualification expression is ignored.
*RST: OFF
TRIGger<m>:QUALify<n>:A:LOGic <Operator>
TRIGger<m>:QUALify<n>:B:LOGic <Operator>
TRIGger<m>:QUALify<n>:C:LOGic <Operator>
TRIGger<m>:QUALify<n>:D:LOGic <Operator>
Defines the logic for the indicated channel:
●
A: CH1
●
B: CH2
●
C: CH3
Command Reference