Suffix:
<m>
.
1..3
1 = A-event only
Parameters:
<ClockEdge> POSitive | NEGative | EITHer
See chapter 16.2.2.3, "Slope Parameter", on page 420.
*RST: POSitive
TRIGger<m>:SPATtern:CSOurce:LEVel <ClockLevel>
Defines the voltage level for the clock signal.
Suffix:
<m>
.
1..3
Indicates the event in a trigger sequence: 1 = A-event, 2 = B-event,
3 = R-event.
Parameters:
<ClockLevel> Range: -10 to 10
Increment: 1E-3
*RST: 0
Default unit: V
TRIGger<m>:SPATtern:PATTern <Pattern>
The pattern contains the bits of the serial data to be found in the data stream. The max-
imum length of the pattern is 128 bit.
Suffix:
<m>
.
1..3
Indicates the event in a trigger sequence: 1 = A-event, 2 = B-event,
3 = R-event.
Parameters:
<Pattern>
Numeric or string pattern, see chapter 16.2.2.6, "Bit Pattern
Parameter", on page 421. The string parameter accepts the bit
value X (don't care).
16.2.6.2 Trigger Qualification
The A-event and B-event in a trigger sequence can have their own trigger qualification.
Qualification is not available for R-events (Event-Suffix m = 3).
The trigger type to which the qualification belongs is defined by a suffix.
Table 16-13: Trigger type suffixes
Suffix Trigger type
1 EDGE
2 GLITch
Command Reference