User manual

Remote Control
R&S
®
RTO
474User Manual 1316.0827.02 ─ 06
Suffix:
<m>
.
1..3
Indicates the event in a trigger sequence: 1 = A-event, 2 = B-event,
3 = R-event.
Parameters:
<ClockLevel> Range: -10 to 10
Increment: 1E-3
*RST: 0
Default unit: V
TRIGger<m>:
DATatoclock:HTIMe <HoldTime>
Sets the minimum time after the clock edge while the data signal must stay steady above
or below the data level.
The hold time can be negative. In this case, the setup time is always positive. The setup/
hold interval starts before the clock edge (setup time) and ends before the clock edge
(hold time). If you change the negative hold time, the setup time is adjusted by the intru-
ment.
Suffix:
<m>
.
1..3
Indicates the event in a trigger sequence: 1 = A-event, 2 = B-event,
3 = R-event.
Parameters:
<HoldTime> Range: -99.999E-9 to 100E-9
Increment: 1E-9
*RST: 0
Default unit: s
TRIGger<m>:DATatoclock:STIMe <SetupTime>
Sets the minimum time before the clock edge while the data signal must stay steady
above or below the data level.
The setup time can be negative. In this case, the hold time is always positive. The setup/
hold interval starts after the clock edge (setup time) and ends after the clock edge (hold
time). If you change the negative setup time, the hold time is adjusted by the intrument.
Suffix:
<m>
.
1..3
Indicates the event in a trigger sequence: 1 = A-event, 2 = B-event,
3 = R-event.
Parameters:
<SetupTime> Range: -99.999E-9 to 100E-9
Increment: 1E-9
*RST: 0
Default unit: s
Command Reference