User manual
Remote Control
R&S
®
RTO
472User Manual 1316.0827.02 ─ 06
Suffix:
<m>
.
1..3
Indicates the event in a trigger sequence: 1 = A-event, 2 = B-event,
3 = R-event.
Parameters:
<RangeMode> INSRange | OUTRange | LTHan | GTHan
INSRange
Triggers on pulses inside a given range. The range is defined by
the slew rate ±delta.
OUTRange
Triggers on pulses outside a given range. The range is defined by
the slew rate ±delta.
LTHan
Triggers on pulses shorter than the given slew rate.
GTHan
Triggers on pulses longer than the given slew rate.
*RST:
GTHan
TRIGger<m>:SLEW:RATE <Time>
For the ranges "Within" and "Outside", the slew rate defines the center of a range which
is defined by the limits "±Delta".
For the ranges "Shorter" and "Longer", the slew rate defines the maximum and minimum
slew rate limits, respectively. When the signal crosses this level, the slew rate measure-
ment starts or stops depending on the selected slope (see TRIGger<m>:SLEW:
SLOPe on page 470).
Suffix:
<m>
.
1..3
Indicates the event in a trigger sequence: 1 = A-event, 2 = B-event,
3 = R-event.
Parameters:
<Time> Range: 100E-12 to 10000
Increment: 100E-9
*RST: 100E-12
Default unit: s
TRIGger<m>:SLEW:DELTa <TimeDelta>
Defines a time range around the slew rate specified using TRIGger<m>:SLEW:RATE.
Suffix:
<m>
.
1..3
Indicates the event in a trigger sequence: 1 = A-event, 2 = B-event,
3 = R-event.