User manual

Remote Control
R&S
®
RTO
471User Manual 1316.0827.02 ─ 06
TRIGger<m>:LEVel<n>:SLEW:UPPer <Level>
Defines the upper voltage threshold. When the signal crosses this level, the slew rate
measurement starts or stops depending on the selected slope.
Suffix:
<m>
.
1..3
Indicates the event in a trigger sequence: 1 = A-event, 2 = B-event,
3 = R-event.
<n> 1..9
Indicates the trigger source:
1...4 = channel 1...4
5...9 = not available
Parameters:
<Level> Range: -10 to 10
Increment: 1E-3
*RST: 0.1
Default unit: V
TRIGger<m>:LEVel<n>:SLEW:LOWer <Level>
Defines the lower voltage threshold. When the signal crosses this level, the slew rate
measurement starts or stops depending on the selected slope.
Suffix:
<m>
.
1..3
Indicates the event in a trigger sequence: 1 = A-event, 2 = B-event,
3 = R-event.
<n> 1..9
Indicates the trigger source:
1...4 = channel 1...4
5...9 = not available
Parameters:
<Level> Range: -10 to 10
Increment: 1E-3
*RST: -0.1
Default unit: V
TRIGger<m>:SLEW:RANGe <RangeMode>
Defines the time limit for the slew rate in relation to the upper or lower trigger level (see
TRIGger<m>:SLEW:RATE on page 472 and TRIGger<m>:SLEW:DELTa
on page 472). The time measurement starts when the signal crosses the first trigger level
- the upper or lower level depending on the selected slope - and stops when the signal
crosses the second level.
Command Reference