468User Manual 1316.0827.02 ─ 06
Parameters:
<WidthDelta> Range: 0 to 432
Increment: 500E-12
*RST: 0
Default unit: s
Timeout Trigger
The timeout trigger is not available for the B-event (Suffix = 2).
TRIGger<m>:TIMeout:RANGe.........................................................................................468
TRIGger<m>:TIMeout:TIME............................................................................................468
TRIGger<m>:TIMeout:RANGe <TimeoutMode>
Defines the relation of the signal level to the trigger level.
Suffix:
<m>
.
1..3
Indicates the event in a trigger sequence: 1 = A-event, 2 = B-event,
3 = R-event.
Parameters:
<TimeoutMode> HIGH | LOW | EITHer
HIGH
The signal level stays above the trigger level.
LOW
The signal level stays below the trigger level.
EITHer
The signal level stays above or below the trigger level.
*RST: HIGH
TRIGger<m>:TIMeout:TIME <Time>
Defines the time limit for the timeout at which the instrument triggers.
Suffix:
<m>
.
1..3
Indicates the event in a trigger sequence: 1 = A-event, 2 = B-event,
3 = R-event.
Parameters:
<Time> Range: 100E-12 to 10000
Increment: 100E-9
*RST: 100E-9
Default unit: s
Interval Trigger
The interval trigger is not available for the B-event (Suffix = 2).
Command Reference