User manual

Remote Control
R&S
®
RTO
465User Manual 1316.0827.02 ─ 06
Parameters:
<WidthDelta> Range: 100E-12 to 864
Increment: 100E-9
*RST: 100E-12
Default unit: s
Window Trigger
The window trigger is not available for the B-event (Suffix = 2).
TRIGger<m>:LEVel<n>:WINDow:UPPer..........................................................................465
TRIGger<m>:LEVel<n>:WINDow:LOWer.........................................................................465
TRIGger<m>:WINDow:RANGe........................................................................................466
TRIGger<m>:WINDow:TIME...........................................................................................466
TRIGger<m>:WINDow:WIDTh.........................................................................................467
TRIGger<m>:WINDow:DELTa.........................................................................................467
TRIGger<m>:LEVel<n>:WINDow:UPPer <Level>
Sets the upper voltage limit for the window.
Suffix:
<m>
.
1..3
Indicates the event in a trigger sequence: 1 = A-event, 2 = B-event,
3 = R-event.
<n> 1..9
Indicates the trigger source:
1...4 = channel 1...4
5...9 = not available
Parameters:
<Level> Range: -10 to 10
Increment: 1E-3
*RST: 0.1
Default unit: V
TRIGger<m>:LEVel<n>:WINDow:LOWer <Level>
Sets the lower voltage limit for the window.
Suffix:
<m>
.
1..3
Indicates the event in a trigger sequence: 1 = A-event, 2 = B-event,
3 = R-event.
<n> 1..9
Indicates the trigger source:
1...4 = channel 1...4
5...9 = not available
Command Reference