User manual
Remote Control
R&S
®
RTO
464User Manual 1316.0827.02 ─ 06
Suffix:
<m>
.
1..3
Indicates the event in a trigger sequence: 1 = A-event, 2 = B-event,
3 = R-event.
Parameters:
<Mode> ANY | LONGer | SHORter | WITHin | OUTSide
ANY
Triggers on all runts fulfilling the level condition, without time lim-
itation.
LONGer
Triggers on runts longer than the given "Runt width".
SHORter
Triggers on runts shorter than the given "Runt width".
WITHin
Triggers if the runt length is inside a given time range. The range
is defined by "Runt width" and "±Delta".
OUTSide
Triggers if the runt length is outside a given time range. The range
is defined by "Runt width" and "±Delta".
*RST:
ANY
TRIGger<m>:RUNT:WIDTh <Width>
Defines the upper or lower voltage threshold. This command is not available if
TRIGger<m>:RUNT:RANGe is set to "ANY".
Suffix:
<m>
.
1..3
Indicates the event in a trigger sequence: 1 = A-event, 2 = B-event,
3 = R-event.
Parameters:
<Width> Range: 100E-12 to 10000
Increment: 100E-9
*RST: 5E-9
Default unit: s
TRIGger<m>:RUNT:DELTa <WidthDelta>
Defines a range around the runt width specified using TRIGger<m>:RUNT:WIDTh. This
command is only available if TRIGger<m>:RUNT:RANGe is set to "WITHin" or "OUT-
Side".
Suffix:
<m>
.
1..3
Indicates the event in a trigger sequence: 1 = A-event, 2 = B-event,
3 = R-event.
Command Reference