User manual
Remote Control
R&S
®
RTO
462User Manual 1316.0827.02 ─ 06
TRIGger<m>:WIDTh:WIDTh <Width>
For the ranges "Within" and "Outside" (defined using TRIGger<m>:WIDTh:RANGe), the
width defines the center of a range which is defined by the limits "±Delta" (see
TRIGger<m>:WIDTh:DELTa on page 462).
For the ranges "Shorter" and "Longer", the width defines the maximum and minimum
pulse width, respectively.
Suffix:
<m>
.
1..3
Indicates the event in a trigger sequence: 1 = A-event, 2 = B-event,
3 = R-event.
Parameters:
<Width> Range: 100E-12 to 10000
Increment: 100E-9
*RST: 5E-9
Default unit: s
TRIGger<m>:WIDTh:DELTa <WidthDelta>
Defines a range around the width value specified using TRIGger<m>:WIDTh:WIDTh.
Suffix:
<m>
.
1..3
Indicates the event in a trigger sequence: 1 = A-event, 2 = B-event,
3 = R-event.
Parameters:
<WidthDelta> Range: 0 to 432
Increment: 500E-12
*RST: 0
Default unit: s
Runt Trigger
The runt trigger is not available for the B-event (Suffix = 2).
TRIGger<m>:RUNT:POLarity..........................................................................................462
TRIGger<m>:LEVel<n>:RUNT:UPPer..............................................................................463
TRIGger<m>:LEVel<n>:RUNT:LOWer.............................................................................463
TRIGger<m>:RUNT:RANGe...........................................................................................463
TRIGger<m>:RUNT:WIDTh............................................................................................464
TRIGger<m>:RUNT:DELTa............................................................................................464
TRIGger<m>:RUNT:POLarity <Polarity>
Suffix:
<m>
.
1..3
Event in a trigger sequence: 1 = A-event, 3 = R-event.
Command Reference