User manual

Remote Control
R&S
®
RTO
460User Manual 1316.0827.02 ─ 06
TRIGger<m>:GLITch:WIDTh...........................................................................................460
TRIGger<m>:GLITch:POLarity <Polarity>
Defines the polarity of a pulse, that is the direction of the first pulse slope.
Suffix:
<m>
.
1..3
Event in a trigger sequence: 1 = A-event, 3 = R-event.
Parameters:
<Polarity> POSitive | NEGative | EITHer
See chapter 16.2.2.4, "Polarity Parameter", on page 420.
*RST: POSitive
TRIGger<m>:
GLITch:RANGe <RangeMode>
Selects which glitches are identified: shorter or longer than the width specified using
TRIGger<m>:GLITch:WIDTh.
Suffix:
<m>
.
1..3
Indicates the event in a trigger sequence: 1 = A-event, 2 = B-event,
3 = R-event.
Parameters:
<RangeMode> SHORter | LONGer
SHORter
Glitches shorter than the specified width are identified.
LONGer
Glitches longer than the specified width are identified.
*RST: SHORter
TRIGger<m>:GLITch:WIDTh <Width>
Sets the length of a glitch. The instrument triggers on pulses shorter or longer than this
value, depnding on the TRIGger<m>:GLITch:RANGe command.
You need to know the expected pulse widths of the circuit to set the glitch width correctly.
Suffix:
<m>
.
1..3
Indicates the event in a trigger sequence: 1 = A-event, 2 = B-event,
3 = R-event.
Parameters:
<Width> Range: 100E-12 to 10000
Increment: 100E-6
*RST: 1E-9
Default unit: s
Command Reference