User manual

Remote Control
R&S
®
RTO
458User Manual 1316.0827.02 ─ 06
Parameters:
<AnalogCutOffHP> KHZ5 | KHZ50 | MHZ50
Cut-off frequency
KHZ5
5 kHz
KHZ50
50 kHz
MHZ50
50 MHz
*RST: KHZ50
TRIGger<m>:ANEDge:CUToff:LOWPass <AnalogCutOffLP>
Frequencies higher than the "Cut-off" frequency are rejected, lower frequencies pass the
filter.
Suffix:
<m>
.
1..3
Only suffix 1 = A-event is allowed, the analog edge trigger is not
available for B- and R-events.
Parameters:
<AnalogCutOffLP> KHZ5 | KHZ50 | MHZ50
KHZ5
5 kHz
KHZ50
50 kHz
MHZ50
50 MHz
*RST: KHZ50
TRIGger<m>:ANEDge:FILTer <Filter>
The analog trigger signal is used for triggering; you can directly select an additional filter
to reject high or low frequencies.
Suffix:
<m>
.
1..3
Only suffix 1 = A-event is allowed, the analog edge trigger is not
available for B- and R-events.