User manual

Remote Control
R&S
®
RTO
457User Manual 1316.0827.02 ─ 06
TRIGger<m>:ANEDge:GND............................................................................................459
TRIGger<m>:ANEDge:SLOPe.........................................................................................459
TRIGger<m>:EDGE:SLOPe <Slope>
Defines the edge for the edge trigger event.
Suffix:
<m>
.
1..3
Event in a trigger sequence: 1 = A-event, 2 = B-event, 3 = R-event.
Parameters:
<Slope> POSitive | NEGative | EITHer
See chapter 16.2.2.3, "Slope Parameter", on page 420.
*RST: POSitive
TRIGger<m>:
ANEDge:COUPling <Coupling>
Sets the coupling for the analog trigger signal.
Suffix:
<m>
.
1..3
Only suffix 1 = A-event is allowed, the analog edge trigger is not
available for B- and R-events.
Parameters:
<Coupling> DC | DCLimit | AC
DC
Direct connection with 50 Ω termination, passes both DC and AC
components of the trigger signal.
DCLimit
Direct connection with 1 MΩ termination, passes both DC and AC
components of the trigger signal.
AC
Connection through DC capacitor, removes DC and very low-fre-
quency components.
*RST: DCLimit
TRIGger<m>:ANEDge:CUToff:HIGHpass <AnalogCutOffHP>
Frequencies below the "Cut-off" frequency are rejected, higher frequencies pass the filter.
Suffix:
<m>
.
1..3
Only suffix 1 = A-event is allowed, the analog edge trigger is not
available for B- and R-events.
Command Reference