User manual
Remote Control
R&S
®
RTO
452User Manual 1316.0827.02 ─ 06
CHANnel<m>:SKEW:MANual <ManualCompensation>
If enabled, the skew offset value (CHANnel<m>:SKEW:TIME) is used for compensation.
This improves horizontal and trigger accuracy.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<ManualCompensation>ON | OFF
*RST: ON
CHANnel<m>:SKEW:TIME <Offset>
Sets an delay value, that is known from the circuit specifics but cannot be compensated
by the instrument automatically. It affects only the selected input channel.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<Offset> Range: -100E-9 to 100E-9
Increment: 1E-12
*RST: 0
Default unit: s
16.2.5.9
Reference (OCXO Option R&S RTO)-B4
SENSe[:ROSCillator]:SOURce........................................................................................452
SENSe[:ROSCillator]:EXTernal:FREQuency.....................................................................452
SENSe[:ROSCillator]:SOURce <RefOscillatorSrc>
Enables the use of the external reference signal instead of the internal OCXO reference.
Parameters:
<RefOscillatorSrc> INTernal | EXTernal
*RST: INTernal
SENSe[:ROSCillator]:EXTernal:FREQuency <ExternalRef>
Sets the frequency of an external reference input signal that is connected to the external
reference input on the rear panel.
Parameters:
<ExternalRef> Range: 1E+6 to 20E+6
Increment: 1E+6
*RST: 10E+6
Default unit: Hz
Command Reference