446User Manual 1316.0827.02 ─ 06
Parameters:
<Mode> RCONtinuous | RSINgle | AUToset | AZERo | SEToffsettomean |
PRINt | SITFile | NOACtion
RCONtinuous
Run continuous: The acquisition is running as long as the probe
button is pressed.
RSINgle
Run single: starts one acquisition.
AUTOSET
Starts the autoset procedure.
AZero
Auto zero: performs an automatic correction of the zero error.
SEToffsettomean
Set offset to mean: performs an automatic compensation for a DC
component of the input signal.
PRINt
Prints the current display according to the printer set with
SYSTem:COMMunicate:PRINter:SELect<1..2>.
SITFile
Save Image To File:
Directs the display image to a file. The MMEMory:NAME command
defines the file name. The file format is defined with HCOPy:
DEVice<m>:LANGuage.
NOACtion
Nothing is started on pressing the micro button.
*RST:
RCONtinuous
PROBe<m>:SETup:TYPE?
Queries the type of the probe.
Suffix:
<m>
.
1..4
Selects the input channel.
Return values:
<Type> String containing one of the following values:
None (no probe detected)
Passive Probe
active single-ended
Usage: Query only
PROBe<m>:SETup:NAME?
Queries the name of the probe.
Command Reference