User manual

Acquisition and Setup
R&S
®
RTO
30User Manual 1316.0827.02 ─ 06
Sample rate
Sets the number of recorded waveform points per second. The sample rate is the recip-
rocal value of the resolution and thus also depends on the acquisition time and the record
length. It considers the samples of the ADC, the additional waveforms points resulting
from resolution enhancement (interpolation and equivalent-time sampling), and the
reduction of waveform points by decimation.
See also:
chapter 2.1.2, "Sampling and Acquisition", on page 15
chapter 2.1.3, "Horizontal System", on page 19
SCPI command:
ACQuire:SRATe on page 432
ADC sample rate
Shows the number of points that are sampled by the ADC in one second. The ADC
sample rate is a constant of the instrument.
SCPI command:
ACQuire:POINts:ARATe on page 432
Resolution
Sets the time between two waveform samples. A fine resolution with low values produces
a more precise waveform record.
SCPI command:
ACQuire:RESolution on page 432
Record length
Indicates the number of waveform samples that build the waveform across the acquisition
time.
SCPI command:
ACQuire:POINts[:VALue] on page 432
Reference for Acquisition and Setup