User manual
Remote Control
R&S
®
RTO
442User Manual 1316.0827.02 ─ 06
PROBe<m>:SERVice:FW:PATH.....................................................................................449
PROBe<m>:SERVice:FW:FLASh....................................................................................450
PROBe<m>:SERVice:STATe..........................................................................................450
PROBe<m>:SETup:STATe?
Queries if the probe at the specified input channel is active (detected) or not active (not
detected). To switch the probe on, use CHANnel<m>:STATe.
Suffix:
<m>
.
1..4
Return values:
<State> DETected | NDETected
*RST: NDETected
Usage: Query only
PROBe<m>:SETup:ATTenuation:MODE <ProbeAttMode>
Set the mode to MANual if the instrument does not detect the probe.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<ProbeAttMode> AUTO | MANual
*RST: AUTO
PROBe<m>:SETup:ATTenuation[:AUTO]?
Queries the attenuation of the probe.
Suffix:
<m>
.
1..4
Selects the input channel.
Return values:
<ProbeAttModeAuto> Range: 0.1 to 1000
Increment: 1
*RST: 1
Default unit: V/V
Usage: Query only
PROBe<m>:SETup:ATTenuation:DEFProbe <SelectedPredefinedProbe>
Selects a predefined current probe. Current probes are not recognized automatically but
the parameters of R&S current probes (R&S RT-ZCxx) are known to the instrument.
Command Reference