<Offset> Negative values move the waveform up, positive values move it
down.
Range: Depends on attenuation factors, input coupling, and
the offset compensation range of active probes. The
nominal offset range for 1:1 attenuation and probe
offset compensation = 0 is specified in the data sheet.
Increment: 0.01
*RST: 0
Default unit: V
CHANnel<m>:BANDwidth <BandwidthLimit>
Selects the bandwidth limit for the indicated channel.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<BandwidthLimit> FULL | B800 | B200 | B20
FULL
Use full bandwidth.
B800
Limit to 800 MHz.
B200
Limit to 200 MHz.
B20
Limit to 20 MHz.
*RST: FULL
CHANnel<m>:IMPedance <Impedance>
Sets the impedance of the channel for power calculations and measurements.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<Impedance> Range: 1 to 100E+3
Increment: 1
*RST: 50
Default unit: Ohm
CHANnel<m>:OVERload <Overload>
Retrieves the overload status of the specified channel from the status bit. When the
overload problem is solved, the command resets the status bit.