User manual
Remote Control
R&S
®
RTO
438User Manual 1316.0827.02 ─ 06
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<Range> Voltage range value
Range: Depends on attenuation factors and coupling. With
1:1 probe and external attenuations and 50 Ω input
coupling, the range is 10 mV to 10 V. For 1 MΩ input
coupling, it is 10 mV to 100 V. If the probe and/or
external attenuation is changed, multiply the range
values by the attenuation factors.
Increment: 0.01
*RST: 0.5
Default unit: V/div
CHANnel<m>:
POSition <Position>
Sets the vertical position of the indicated channel as a graphical value.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<Position> Positive values move the waveform up, negative values move it
down.
Range: -5 to 5
Increment: 0.02
*RST: 0
Default unit: div
CHANnel<m>:OFFSet <Offset>
The offset voltage is subtracted to correct an offset-affected signal. The offset of a signal
is determined and set by the autoset procedure.
See also: "Offset" on page 39
Suffix:
<m>
.
1..4
Selects the input channel.
Command Reference