User manual
Remote Control
R&S
®
RTO
437User Manual 1316.0827.02 ─ 06
Parameters:
<Coupling> DC | DCLimit | AC
DC
Direct connection with 50 Ωtermination.
DCLimit
Direct connection with 1 MΩ termination.
AC
Connection through DC capacitor.
*RST: DCLimit
CHANnel<m>:
GND <State>
Connects the signal to the ground.
Suffix:
<m>
.
1..4
Parameters:
<State> ON | OFF
*RST: OFF
CHANnel<m>:SCALe <Scale>
Sets the vertical scale for the indicated channel.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<Scale> Scale value, given in Volts per division.
Range: Depends on attenuation factors and coupling. With
1:1 probe and external attenuations and 50 Ω input
coupling, the vertical scale (input sensitivity) is 1 mV/
div to 1 V/div. For 1 MΩ input coupling, it is 1 mV/div
to 10 V/div. If the probe and/or external attenuation
is changed, multiply the values by the attenuation
factors to get the actual scale range.
Increment: 1E-3
*RST: 0.05
Default unit: V/div
CHANnel<m>:RANGe <Range>
Sets the voltage range across the 10 vertical divisions of the diagram. Use the command
alternativly instead of CHANnel<m>:SCALe.
Command Reference