User manual
Remote Control
R&S
®
RTO
433User Manual 1316.0827.02 ─ 06
ACQuire:MODE <EnhancementMode>
Selects the method of adding waveform points to the samples of the ADC in order to fill
the record length.
See also: "Resolution enhancement" on page 32.
Parameters:
<EnhancementMode>RTIMe | ITIMe | ETIMe
RTIMe
Real Time Mode: The sampled points of the input signal are used
to build the waveform, no waveform points are added.
ITIMe
Interpolated time: Interpolation of waveform points with the
method set by the interpolation mode, see ACQuire:
INTerpolate on page 433.
ETIMe
Equivalent time: The waveform points are taken from several
acquisitions of a repetive signal at a different time in relation to the
trigger point.
*RST:
ITIMe
ACQuire:INTerpolate <IntpolMode>
Selects the interpolation method if ACQUire:MODE ITIMe (interpolated time) is set for
enhancement.
See also: "Interpolation mode" on page 33.
Parameters:
<IntpolMode> LINear | SINX | SMHD
LINear
Linear interpolation between two adjacent sample points
SINX
Interpolation by means of a sin(x)/x curve.
SMHD
Sample/Hold causes a histogram-like interpolation.
*RST: SINX
CHANnel<m>[:WAVeform<n>][:STATe] <State>
Activates or deactivates a waveform. [:STATe] can be omitted.
Suffix:
<m>
.
1..4
Selects the input channel.
Command Reference