ACQuire:POINts:ARATe?
Retrieves the sample rate of the ADC, that is the number of points that are sampled by
the ADC in one second.
Return values:
<ADCSampleRate> Range: 10 to 20E+9
Increment: 1
*RST: 10E+9
Default unit: Hz
Usage: Query only
ACQuire:
SRATe <SampleRate>
Defines the sample rate, that is the number of recorded waveform samples per second.
See also: "Sample rate" on page 30.
Parameters:
<SampleRate> Range: 2 to 20E+12
Increment: 1
*RST: 20E+3
Default unit: Sa/s
ACQuire:RESolution <Resolution>
Indicates the time between two waveform points in the record.
Parameters:
<Resolution> A fine resolution with low values produces a more precise wave-
form record.
Range: 1E-15 to 0.5
Increment: 10E-12
*RST: 500E-6
Default unit: s
ACQuire:POINts[:VALue] <RecordLength>
Indicates the record length, the number of recorded waveform points that build the wave-
form across the acquisition time. [:VALue] can be omitted.
Parameters:
<RecordLength> Number of recorded waveform points.
Range: 1000 to 1000000000
Increment: 2
*RST: 1000
Default unit: Sa
Command Reference