User manual

Remote Control
R&S
®
RTO
429User Manual 1316.0827.02 ─ 06
TIMebase:SCALe <TimeScale>
Sets the horizontal scale - the time per division on the x-axis - for all channel and math
waveforms.
The setting accuracy depends on the current resolution (sample rate) and the setting for
resolution enhancement:
In interpolated time mode if sample rate > ADC sample rate:
Any value for the horizontal scale can be set due to the interpolation factor.
In real time mode and equivalent time mode for all sample rates; and in interpolated
time mode if sample rate < ADC sample rate:
The resolution is an integer multiple of the ADC sample rate.
Parameters:
<TimeScale> Range: 25E-12 to 50
Increment: 1E-12
*RST: 10E-9
Default unit: s/div
TIMebase:RANGe <AcquisitionTime>
Defines the time of one acquisition, that is the time across the 10 divisions of the diagram:
TimeScale*10.
Parameters:
<AcquisitionTime> Range: 250E-12 to 500
Increment: 1E-12
*RST: 0.5
Default unit: s
TIMebase:DIVisions?
Queries the number of horizontal divisions on the screen. The number cannot be
changed.
Return values:
<HorizDivCount> Range: 4 to 20
Increment: 2
*RST: 10
Usage: Query only
TIMebase:POSition <Offset>
Defines the trigger offset - the time interval between trigger point and reference point to
analize the signal some time before or after the trigger event.
See also: TIMebase:REFerence on page 430
Command Reference