User manual
Remote Control
R&S
®
RTO
420User Manual 1316.0827.02 ─ 06
Depending on the command, not all parameter values are allowed for usage:
●
SEARch, EXPort commands, XY-diagram: Only analog channel waveforms (CxWy),
math (Mx) and reference waveforms (Rx) are allowed.
●
Sources for serial buses: Only waveform 1 of analog channels (CxW1), math (Mx)
and reference waveforms (Rx) are allowed.
Waveform Description
C1W1 | C1W2 | C1W3 Channel 1 waveforms
C2W1 | C2W2 | C2W3 Channel 2waveforms
C2W3 | C3W1 | C3W2 Channel 3 waveforms
C4W1 | C4W2 | C4W3 Channel 4 waveforms
M1 | M2 | M3 | M4 Math waveforms
R1 | R2 | R3 | R4 Reference waveforms
XY1 | XY2 | XY3 | XY4 XY-waveforms
MRESult1 | MRESult2 | MRESult3 | MRESult4
| MRESult5 | MRESult6 | MRESult7 | MRESult8
IMResult
Measurement results
Result of immediate measurements available on Tektronix
instruments. Only relevant for Tektronix emulation (option
R&S RTO-K301).
SBUS1 | SBUS2 | SBUS3 | SBUS4 Serial buses
D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7 | D8 | D9 |
D10 | D11 | D12 | D13 | D14 | D15
Digital channels (option R&S RTO-B1)
MSOB1 | MSOB2 | MSOB3 | MSOB4
Digital buses (option R&S RTO-B1)
16.2.2.3 Slope Parameter
The slope parameter is used with several trigger and search condition commands.
Slope Description
POSitive Rising edge, that is a positive voltage change.
NEGative Falling edge, that is a negative voltage change
EITHer rising as well as the falling edge.
16.2.2.4 Polarity Parameter
The polarity parameter is used with several trigger and search condition commands.
Command Reference