Remote Control
R&S
®
RTO
402User Manual 1316.0827.02 ─ 06
a <New Line> with EOI
an EOI together with the last data byte
Several commands in a command line must be separated by a semicolon ";". If the next
command belongs to a different command system, the semicolon is followed by a colon.
Example:
MMEM:COPY "Test1","MeasurementXY";:HCOP:ITEM ALL
This command line contains two commands. The first command belongs to the MMEM
system, the second command belongs to the HCOP system.
If the successive commands belong to the same system, having one or several levels in
common, the command line can be abbreviated. To this end, the second command after
the semicolon starts with the level that lies below the common levels. The colon following
the semicolon must be omitted in this case.
Example:
HCOP:ITEM ALL;:HCOP:IMM
This command line contains two commands. Both commands are part of the HCOP com-
mand system, i.e. they have one level in common.
When abbreviating the command line, the second command begins with the level below
HCOP. The colon after the semicolon is omitted. The abbreviated form of the command
line reads as follows:
HCOP:ITEM ALL;IMM
A new command line always begins with the complete path.
Example:
HCOP:ITEM ALL
HCOP:IMM
16.1.4.6 Responses to Queries
A query is defined for each setting command unless explicitly specified otherwise. It is
formed by adding a question mark to the associated setting command. According to
SCPI, the responses to queries are partly subject to stricter rules than in standard IEEE
488.2.
The requested parameter is transmitted without a header.
Example: HCOP:PAGE:ORI?, Response: LAND
Maximum values, minimum values and all other quantities that are requested via a
special text parameter are returned as numeric values.
Example: SENSe:FREQuency:STOP? MAX, Response: 3.5E9
Numeric values are output without a unit. Physical quantities are referred to the basic
units or to the units set using the Unit command. The response 3.5E9 in the pre-
vious example stands for 3.5 GHz.
Truth values (Boolean values) are returned as 0 (for OFF) and 1 (for ON).