User manual
General Instrument Setup
R&S
®
RTO
367User Manual 1316.0827.02 ─ 06
"FLOAT"
Binary data values are written to a file.
SCPI command:
FORMat[:DATA] on page 426
Bit pattern format
Sets the format for all bit pattern queries.
SCPI command:
FORMat:BPATtern on page 427
SCPI emulation mode
If option R&S RTO-K301 is installed, you can define the remote control behavior of the
instrument.
"Tektronix
DPO7000 /
TDS540"
Emulated remote commands of these Tektronix oscilloscopes are used.
"Native"
By default, remote commands of R&S RTO are used.
SCPI command:
SYSTem:LANGuage on page 427
SCPI emulation mode settings
"IDN response" and "OPT response" define the responses to commands IDN*? and
OPT*? which are expected by the remote control scripts. Instead of the actual RTO iden-
tification and options, these specified strings are returned. Use the "Description" field to
add a comment to the response settings.
The settings are only relevant if the SCPI emulation mode is set to Tektronix emulation
(requires option R&S RTO-K301).
13.2.1.6 LXI
This tab provides settings for LXI, which allows you to connect your R&S RTO to other
devices in a network. For details, see chapter 14.4.2, "LXI Configuration", on page 379.
Reference for General Instrument Settings