4. Tap "Load" to load the settings from the specified file.
Tap "Open" to load the settings from a different file. Select the file from the file selec-
tion dialog box.
The saved settings are loaded to the R&S RTO.
12.1.4 Restoring Settings
When you have changed many different settings on the instrument and are no longer
sure which settings are causing which effect in the measurement, you may want to restore
the default settings and start anew. Depending on the situation and which data is to be
restored, the following methods are available:
●
Restoring the instrument settings to their default values
● Restoring settings from a file (see "To load settings from a file" on page 347)
●
Restoring the default instrument settings and user-specific settings to a saved state
in one step during one measurement session
● Restoring all settings on the R&S RTO to the factory-defined values
To restore the instrument settings to their default values
► Press the PRESET key.
The instrument settings are restored to their default values.
To restore the default instrument settings and user-specific settings to a saved
state in one step
This method is only available during one measurement session.
1. Press the FILE key.
2. Select the "User-defined Preset" tab.
3. Tap the "Create SaveSet" button to save the current user-specific settings temporarily
in a SaveSet. The SaveSet remains available until you switch off the instrument, then
it is deleted.
4. Enable the "Enable user-defined preset" option.
5. At any time during the same measurement session, press the PRESET key.
The instrument settings are reset to their default values, the user-specific settings are
reset to the values saved in the SaveSet.
To restore all settings on the R&S RTO to the factory-defined values
► Tap the "Factory Reset" button.
All settings on the R&S RTO are reset to their factory-defined values.