Mixed Signal Option (MSO, R&S RTO-B1)
R&S
®
RTO
333User Manual 1316.0827.02 ─ 06
Make sure that the trigger sequence is set to "A only".
Additionally, you can define trigger holdoff conditions in the "Sequence" tab. See also:
"Holdoff mode" on page 81.
Source
If the Mixed Signal Option is installed, the variety of trigger sources of the A-event setup
is enhanced with specific digital trigger sources. You can select as trigger source:
one of the digital channels "D0" ... "D15"
a logic combination of digital channels: "Logic"
one of the parallel buses "Par. bus1" ... "Par. bus4"
SCPI command:
TRIGger<m>:SOURce on page 453
Type
Depending on the selected source, the instrument provides the appropriate trigger types
and the corresponding trigger settings. For mixed signal analysis, the following trigger
types are available:
Edge, see page 333
Width, see page 334
Timeout, see page 336
Data2Clock, see page 337
State, see page 338
Pattern, see page 339
Serial Pattern, see page 341
SCPI command:
TRIGger<m>:PARallel:TYPE on page 743
11.3.4.2 Edge
Using the edge trigger, you can also trigger on a single digital channel (a logic bit), and
a logical combination of digital channels.
Depending on the selected trigger source, different trigger settings are available. The
trigger level is already set - in MSO the logical threshold is used as trigger level.
Reference for MSO