297User Manual 1316.0827.02 ─ 06
Fig. 10-12: Structure of a byte field
Data bytes are transmitted LSB first.
The identifier byte consists of 6 bits for the frame identifier and two parity bits. This com-
bination is known as protected identifier.
Trigger
The R&S RTO can trigger on various parts of LIN frames. The data line must be con-
nected to an input channel, triggering on math and reference waveforms is not possible.
You can trigger on:
●
Frame start (synchronization field)
●
Specific slave identifier or identifier range
●
Data pattern in the message
●
Wake up signal
●
Checksum error (error in data), parity error (error in identifier)
10.6.2 Reference for LIN
10.6.2.1 LIN Configuration
Make sure that the tab of the correct serial bus is selected on the left side.
LIN (Option R&S RTO-K3)