User manual
Protocol Analysis
R&S
®
RTO
283User Manual 1316.0827.02 ─ 06
Trigger source
Selects the transmitter or receiver line as trigger source.
SCPI command:
TRIGger<m>:UART:SOURce on page 680
Data conditions
Specify the data conditions if the trigger type is set to "Data".
Condition ← Data conditions
Selects the operator for the "Data" pattern: "Equal" or "Not equal".
SCPI command:
TRIGger<m>:UART:FCONdition on page 681
Pattern ← Data conditions
Specifies the data pattern to be found on the specified trigger source, in binary or hex
format. Enter the words in msb first bit order. The starting point of the pattern is defined
by Position and Index min, Index max.
The bit pattern editor helps you to enter the pattern, see chapter 10.1.4, "Bit Pattern
Editor", on page 255.
SCPI command:
TRIGger<m>:UART:DATA on page 681
Position ← Data conditions
Operator for the data position. You can defined an exact position, or a position range.
SCPI command:
TRIGger<m>:UART:DPOPerator on page 680
Index min, Index max ← Data conditions
Sets the number of words before the first word of interest. These offset words are ignored.
If the Position defines a range, the first and the last word of interest are defined.
SCPI command:
TRIGger<m>:UART:DPOSition on page 681
TRIGger<m>:UART:DPTO on page 681
10.4.2.3 UART Decode Results (Option R&S RTO-K2)
If the option is installed, the "Decode" function in the "Configuration" tab is available.
Enable "Decode" to display the decoded signal below the waveforms.
Additionally, you can display the binary signal and the detailed decoding results, see
chapter 10.1.2, "Display", on page 251.
UART / RS232