User manual

Protocol Analysis
R&S
®
RTO
282User Manual 1316.0827.02 ─ 06
10.4.2.2 UART Trigger
The "Events" tab of the "Trigger" dialog box provides the trigger settings for the configured
serial buses.
Make sure that:
the trigger sequence is set to "A only"
the trigger source is "Serial bus", and the data source(s) of the bus are channel signals
the correct serial bus is selected
the correct protocol is selected
Type
Selects the trigger type for UART analysis.
"Start bit"
Triggers on a start bit. The start bit is the first low bit after a stop bit.
"Packet start"
Triggers on the begin of a data packet. The frame start is configured
with "Packets" on page 281.
"Data"
Trigger on a serial pattern at a defined position in the data packet. The
pattern can include several subsequent symbols (data frames).
See "Data conditions" on page 283.
"Parity error"
Triggers on a parity error indicating a transmission error. This trigger
type is only available if a parity is configured for the UART bus.
"Break condi-
tion"
Triggers if a start bit is not followed by a stop bit, the data line remains
at logic 0 for longer than a UART word.
"Stop error"
Triggers if the stop bit is a logic 0.
SCPI command:
TRIGger<m>:UART:TYPE on page 680
UART / RS232