User manual
Protocol Analysis
R&S
®
RTO
265User Manual 1316.0827.02 ─ 06
Data setup
Specifies the data conditions:
Position
← Data setup
Operator for the data position within a frame. You can define an exact position, or a
position range. Select "Any", if the position of the required pattern is not relevant.
SCPI command:
TRIGger<m>:I2C:DPOPerator on page 652
Index min, Index max ← Data setup
Sets the number of data bytes to be skipped after the address. The index 0 is associated
with the first data byte. If the Position defines a range, the first and the last byte of interest
are defined.
SCPI command:
TRIGger<m>:I2C:DPOSition on page 653
TRIGger<m>:I2C:DPTO on page 653
Condition ← Data setup
Selects the operator for the "Data" pattern: "Equal", "Not equal", or a range definition.
SCPI command:
TRIGger<m>:I2C:DCONdition on page 653
Value min / Value max ← Data setup
Specifies the data bit pattern. Enter the bytes in msb first bit order. The maximum pattern
length is 64 bit. Waveform data is compared with the pattern byte-by-byte.
The instrument ensures that the max value is always ≥ the min value, and X bits (don't
care) are at the same position in both values.
The bit pattern editor helps you to enter the pattern, see chapter 10.1.4, "Bit Pattern
Editor", on page 255.
SCPI command:
TRIGger<m>:I2C:DMIN on page 653
TRIGger<m>:I2C:DMAX on page 654
10.2.3.3 I²C Decode Results (Option R&S RTO-K1)
If the option is installed, the "Decode" function in the "Configuration" tab is available.
Enable "Decode" to display the decoded signal below the waveforms.
I²C