User manual
Search Functions
R&S
®
RTO
243User Manual 1316.0827.02 ─ 06
Clear results
Clears the search results once to start a new search.
SCPI command:
SEARch:CLEar on page 612
Search conditions
Defines the search conditions, depending on the search type (spectrum or trigger). For
trigger searches, most parameters available for trigger event definition can also be con-
figured as search conditions. However, not only signal channels, but also math and ref-
erence waveforms can be selected as the search source.
For details on trigger event definition, see chapter 3.3.1, "Events", on page 58.
For frequency marker searches, only the peak excursion can be defined as a parameter
for the search.
The following search conditions are available for trigger searches:
● Edge, see page 62
● Glitch, see page 63
● Width, see page 64
● Runt, see page 66
● Window, see page 67
● Timeout, see page 69
● Interval, see page 70
● Slew Rate, see page 71
● Data2Clock, see page 72
Note: Serial patterns are currently not supported as search conditions.
While the interval and width triggers can only analyze either positive or negative polarity,
searching for an interval or width is also possible for both polarities at the same time
("Either").
The following general functions are available in addition to the trigger-related settings, or
alternatively for frequency-related searches.
SCPI command:
The remote control commands are described in chapter 16.2.13.3, "Search Condi-
tions", on page 613.
Reference for Search Settings