Mask Testing
R&S
®
RTO
229User Manual 1316.0827.02 ─ 06
Insert a new point before the selected point.
Append a new point at the end of the list.
Remove the selected point from the list.
Change the x- and y-values of each point. To scale or move the complete segment,
use offset and factor values, see Rescale.
SCPI command:
MTESt:SEGMent:POINt:ADD on page 602
MTESt:SEGMent:POINt:REMove on page 602
MTESt:SEGMent:POINt:INSert on page 602
MTESt:SEGMent:POINt:X on page 603
MTESt:SEGMent:POINt:Y on page 603
MTESt:SEGMent:POINt:COUNt on page 603
Rescale
You can rescale and move mask segments by numerical input of factors and offsets. The
values change the selected mask segment and take effect on "Recalculate".
Offset X ← Rescale
Moves the mask segment horizontally. The specified offset is added to the x-values of all
points of the selected mask segment.
To take effect, tap "Recalculate".
SCPI command:
MTESt:SEGMent:RESCale:XOFFset on page 605
Factor X ← Rescale
Stretches or compresses the selected mask segment in horizontal direction. The x-values
of all points of the selected mask segment are multiplied with this factor. Factors >1
stretch the mask segment, while factors between 0 and 1 compress it. Negative values
are possible and change the algebraic sign.
To take effect, tap "Recalculate".
SCPI command:
MTESt:SEGMent:RESCale:XFACtor on page 604
Offset Y ← Rescale
Moves the mask segment vertically. The specified offset is added to the y-values of all
points of the selected mask segment.
To take effect, tap "Recalculate".
SCPI command:
MTESt:SEGMent:RESCale:YOFFset on page 605
Factor Y ← Rescale
Stretches or compresses the selected mask segment in vertical direction. The y-values
of all points of the selected mask segment are multiplied with this factor. Factors >1
stretch the mask segment, while factors between 0 and 1 compress it. Negative values
are possible and change the algebraic sign.
Reference for Masks