User manual

Mask Testing
R&S
®
RTO
225User Manual 1316.0827.02 ─ 06
The content of the "Test Definition" tab depends on the selected definition type: "User"
or "Waveform". If "Waveform" is selected, the main mask settings can be set directly on
the "Test Definition" tab. For a description of these settings, see chapter 8.3.2.2, "Mask
Definition: Waveform Mask", on page 230.
Make sure that the correct "Mask Test" tab is selected on the left side before you enter
the settings.
SCPI commands:
MTESt:ADD on page 597
MTESt:REMove on page 598
Enable test
Activates and deactivates the mask test. If the acquisition is running, the test starts
immediately. Otherwise, the test starts when acquisition is started.
The testing is stopped when acquisition is stopped, or if a stop action is configured with
Stop acq..
Closing the result box also disables the mask test.
SCPI command:
MTESt[:STATe] on page 598
Source
Selects the waveform to be tested against the mask. All channel waveforms can be tes-
ted.
SCPI command:
MTESt:SOURce on page 598