User manual

Mask Testing
R&S
®
RTO
224User Manual 1316.0827.02 ─ 06
8.2.6 Mask Testing on History Acquisitions
In the same way as for running acquisitions, you can set up and perform the mask testing
also on history waveforms.
The requirements for mask testing on history waveforms are also the same:
The mask is defined, see chapter 8.2.1.1, "Creating New User
Masks", on page 217 and chapter 8.2.1.2, "Modifying User Masks", on page 218.
The mask test is defined, see chapter 8.2.2, "Setting Up a Mask
Test", on page 221
The mask display is configured, see chapter 8.2.3, "Configuring the Mask and Hit
Display", on page 221.
1. Perform and finish the acquisition.
2. Press HISTORY.
3. In the quick-access "History" dialog box, tap "Play".
The mask testing is performed on the complete history memory, starting with the
oldest acquisition. The state of the mask test is set to "Finished" when "Nx Single
count" acquisitions are tested.
For details on history, see chapter 4.4, "History", on page 114.
8.3 Reference for Masks
8.3.1 Test Definition
The "Test Definition" tab provides all settings for the mask test itself: the waveform to be
tested, pass/fail conditions, and saving/loading the mask definition.
Reference for Masks