User manual
Mask Testing
R&S
®
RTO
222User Manual 1316.0827.02 ─ 06
3. Select "Show mask" to display the masks of all enabled mask tests on the screen.
4. Define how the sample hits are displayed:
a) Select "Highlight hits" to display the sample hits.
b) Set the "Highlight time" or "Infinite highlight".
Set the "Color" of the sample hits.
5. Define the color of the masks segments depending on the violation state:
● Mask without violation
● Mask with violation
● Mask with contact: This color shows that the edge of the mask segment was
touched. In this case, the resolution is not sufficient to detect if the mask was
really hit or not. Zoom into the concerned area to see the correct result.
8.2.4 Running a Mask Test
Before you can start a mask test, make sure that the mask setup is complete:
● The mask is defined, see chapter 8.2.1.1, "Creating New User
Masks", on page 217 and chapter 8.2.1.2, "Modifying User Masks", on page 218.
● The mask test is defined, see chapter 8.2.2, "Setting Up a Mask
Test", on page 221
Working with Masks