User manual

Mask Testing
R&S
®
RTO
216User Manual 1316.0827.02 ─ 06
Number of tested acquisitions.
SCPI command:
MTESt:RESult:COUNt:WAVeforms on page 609
Acq. remaining
Remaining acquisitions until "Average count / Nx Single count" is reached.
The value is useful if you test a specified number of acquisitions with action "Stop acqui-
sition" on violation, or if the acquisition has been stopped manually before the required
number of acquisitions has been acquired.
See also: chapter 8.2.4, "Running a Mask Test", on page 222.
SCPI command:
MTESt:RESult:COUNt:REMaining on page 609
State
Shows if the test has been completed. The state is set to "Finished" when "Nx Single
count" acquisitions are tested and the number of "Acq. remaining" is 0. as long as the
number of tested acquisitions is less the "Nx Single count" number, the state is "Run-
ning".
If you run the acquisition with RUN CONT, or the number of played history acquisitions
exceeds "Nx Single count", the mask testing is perfomed according to fail criteria settings
independently of the test state. The testing is not stopped when the state is set to "Fin-
ished".
SCPI command:
MTESt:RESult:STATe on page 609
Sample hits
Number of samples that hit the mask.
SCPI command:
MTESt:RESult:COUNt:FAILures on page 610
Acquisition hits
Number of acquisitions that contained at least one sample hit.
SCPI command:
MTESt:RESult:COUNt:FWAVeforms on page 610
Fail rate
Ratio of acquisition hits to the number of tested acquisitions.
SCPI command:
MTESt:RESult:FRATe on page 610
About Mask Testing