User manual

Measurements
R&S
®
RTO
177User Manual 1316.0827.02 ─ 06
Beep
Generates a beep sound.
"No function"
The action is not initiated.
"On violation"
The action is initiated if the limits or margins are exceeded during the
measurement.
"On successful
completion"
The action is initiated if the limits or margins were not exceeded during
the entire measurement.
SCPI command:
MEASurement<m>:ONViolation:BEEP on page 574
Stop acq
Stops data acquisition on violation.
SCPI command:
MEASurement<m>:ONViolation:ACQStop on page 574
Print
Prints a screenshot including the measurement results to the printer defined in the
"Print" dialog box (see chapter 12.1.1, "Configuring Printer Output and Print-
ing", on page 344).
SCPI command:
MEASurement<m>:ONViolation:PRINt on page 574
Save Wfm
Saves the waveform data to the file specified in FILE > "Save/Recall" > "Waveform".
SCPI command:
MEASurement<m>:ONViolation:WFMSave on page 575
Visible histogram
Shows the number of currently shown histograms.
5.2.3.10 Measurements - Sensors Tab
Environment sensors can provide additional information during a measurement, e.g. the
temperature. The sensor results are displayed as a background color according to the
selected color table in the measurement diagram. Thus, the influence of temperature or
humidity changes on the measurement results is visible directly.
The "Sensors" tab contains the settings for the display of sensor measurement results
and the currently measured value. Each connected sensor has its own settings tab. For
each sensor, two tabs are available to set up the display of temperature and humidity
results.
Automatic Measurements