User manual
Measurements
R&S
®
RTO
176User Manual 1316.0827.02 ─ 06
Reset count
Defines the number of measurements after which the statistics are reset.
SCPI command:
MEASurement<m>:STATistics:RCOunt on page 567
Statistics average count
Defines the number of measurements for which the statistical average is calculated.
SCPI command:
MEASurement<m>:STATistics:WEIGht on page 569
Total measurement time
Defines the total duration of the long term measurement.
This setting is only available if "Reset statistics mode" is set to "Time".
SCPI command:
MEASurement<m>:LTMeas:TIME on page 569
Measurement points
Defines the total number of points to be measured during the long term measurement.
SCPI command:
MEASurement<m>:LTMeas:COUNt on page 569
5.2.3.9 Measurements - Event Actions Tab
The settings in this tab define what happens when the limits and margins defined in the
"Setup" tab are exceeded if limit checking is enabled. Independant of these settings, an
icon is displayed in the result box, see chapter 5.2.1, "Measurement Types and
Results", on page 129.
Note that the violation actions do not distinguish between a margin violation and a limit
violation. However, different icons are displayed in the result box.
For each action, you can define the event on which the action is initiated: on violation of
margin or limits, or on successful completion without any violation.
Meas 1/2/3/4/5/6/7/8
Selects one of the eight available measurements.
Automatic Measurements