User manual
Measurements
R&S
®
RTO
173User Manual 1316.0827.02 ─ 06
Multiple measurement
Performs multiple measurements on the same source waveform, e.g. measures the rise
time for all pulses in the waveform, not only the first. This is useful when calculating
statistics; however, it reduces the performance of the instrument.
SCPI command:
MEASurement<m>:MULTiple on page 527
Limit
Sets the maximum number of measurements per acquisition if "Multiple measurement"
is enabled.
SCPI command:
MEASurement<m>:MNOMeas on page 527
Display result lines
Displays intermediate result lines in the measurement waveform (e.g. signal thresholds)
required to obtain the measurement result.
SCPI command:
MEASurement<m>:DISPlay:RESults on page 564
Display reference levels
Displays the reference levels used for the measurement in the diagram.
SCPI command:
MEASurement<m>:DISPlay:LEVels on page 564
Show result control panel
Extends the result box of the current measurement with the source settings and the sta-
tistics enabling. Thus you can check and change the measurement sources directly in
the results box, and also enable statistics there.
Clear Results
Clears the measurement results to begin a new measurement.
5.2.3.8 Measurements - Long Term/Statistics Tab
The settings in this tab allow you to configure long term measurements, including statis-
tics over a longer period of time.
Automatic Measurements