User manual
Measurements
R&S
®
RTO
155User Manual 1316.0827.02 ─ 06
●
5/50/95
●
10/50/90
●
20/50/80
●
User defined: Enter "Upper ref level", "Middle ref level", and "Lower ref level".
For example, for "5/50/95":
●
lower reference level = 5% of high signal level
●
middle reference level = 50% of high signal level
●
upper reference level = 95% of high signal level
SCPI command:
REFLevel<m>:RELative:MODE on page 529
Upper ref level, Middle ref level, Lower ref level
Define any reference levels in percent, if "Relative levels" is set to "User-defined".
SCPI command:
REFLevel<m>:RELative:UPPer on page 535
REFLevel<m>:RELative:MIDDle on page 536
REFLevel<m>:RELative:LOWer on page 536
High
The signal value that represents a high level - for manual reference level mode, absolute
level definition and user signal level.
SCPI command:
REFLevel<m>:ABSolute:HIGH on page 532
MEASurement<m>:REFLevel:RESult:SIGHigh on page 540
Low
The signal value that represents a low level - for manual reference level mode, absolute
level definition and user signal level.
SCPI command:
REFLevel<m>:ABSolute:LOW on page 533
MEASurement<m>:REFLevel:RESult:SIGLow on page 540
Top distance
The distance between the high signal level and the upper reference level - for manual
reference level mode and absolute level definition.
SCPI command:
REFLevel<m>:ABSolute:TDIStance on page 533
Bottom distance
The distance between the lower reference level and the low signal value - for manual
reference level mode and absolute level definition.
SCPI command:
REFLevel<m>:ABSolute:BDIStance on page 533
Automatic Measurements