Measurements
R&S
®
RTO
126User Manual 1316.0827.02 ─ 06
"Horizontal cursors" The horizontal cursors are positioned along the waveform or can
be positioned manually.
"Vertical cursors" The vertical cursors are positioned manually.
"Both horizontal and vertical cursors" The horizontal cursors are positioned along the waveform or can
be positioned manually.
The vertical cursors are positioned manually.
Y user position 1/2
Defines the position of the horizontal cursors in the time domain.
SCPI command:
CURSor<m>:Y1Position on page 518
CURSor<m>:Y2Position on page 518
Track waveform
The horizontal cursors track the waveform, i.e. cursor 1 indicates the current maximum,
cursor 2 indicates the current minimum. If the waveform changes, e.g. during a running
measurement, the cursors move along with it. If both horizontal and vertical cursors are
displayed, the horizontal cursors are positioned to the crossing points of the vertical cur-
sors with the waveform.
SCPI command:
CURSor<m>:TRACking[:STATe] on page 516
Coupling
Couples the cursors of a set so that the distance between the two remains the same if
one cursor is moved.
X position 1/2
Defines the position of the vertical cursors.
Envelope wfm selection 1/2
If the waveform arithmetics are set to envelope waveform (see "Wfm Arith-
metic" on page 34) and "Track waveform" is enabled, this setting defines which horizontal
cursor is positioned to the maximum and which to the the minimum envelope values.
These settings are only available if both horizontal and vertical cursors are enabled.
"Minimum"
The horizontal cursor is set to the crossing point of the vertical cursor
with the minimum waveform envelope.
"Maximum"
The horizontal cursor is set to the crossing point of the vertical cursor
with the maximum waveform envelope.
SCPI command:
CURSor<m>:X1ENvelope on page 519
CURSor<m>:X2ENvelope on page 519
All Off
Disables all cursor measurements at once.
Cursor measurements