User manual
Display
R&S
®
RTO
101User Manual 1316.0827.02 ─ 06
Tool Settings
Defines the visibility of selected toolbar icons.
Show All
Displays all available toolbar icons.
Hide All
Hides all toolbar icons.
Show date/time
Displays the current date and time on the toolbar.
4.1.3.5 Performance
Information on the current acquisition performance values of the R&S RTO is available
by selecting the "Display > Performance" menu entry.
The instrument groups acquired waveforms together in a frame, and displays the frame
content. The maximum number of frames displayed per second is about 30. The current
number of frames per second is indicated as recoprocal "Time per frame". If the time
scale decreases, and thus the number of Acquisitions per second also decreases, the
number od acquisisions per frame can drop to 1.
4.1.3.6 Clear Screen Results
The function "Clear screen results" in the "Display" menu resets all results in all mea-
surement result boxes including long term measurement and statistic results and deletes
the current measurement waveforms.
4.2 Zoom
The zoom functions allow you to magnify a specific section of the diagram in order to view
more details. You can define several zoom areas for the same diagram and even couple
them, or you use the hardware zoom.
4.2.1 Methods of Zooming
The R&S RTO provides the usual way of zooming: You define the section of a diagram
that you want to magnify, and the zoomed view is shown in a separate zoom diagram.
Additionally, you can magnify the diagram directly: The hardware zoom changes the
horizontal and vertical scales of the diagram so that you see the selected section.
Zoom