Specifications

Diamondback360®CoronaryOrbitalAtherectomySystem
41
Issue
number
Issue Solution
10 The low saline
level sensor (red
LED) is illuminated
Note: The OAS pump will stop
pumping saline and supplying power
to the OAD 30 seconds after the low
saline level sensor activates while the
OAD is spinning.
1. If there is less than 200 mL of
saline left in the bag of saline and
VIPERSLIDE Lubricant, replace
the bag with a new 1000 mL bag
of normal saline and VIPERSLIDE
Lubricant solution.
2. Ensure that the bag of saline and
VIPERSLIDE Lubricant is hanging
freely from the saline bag open
hook and that the low saline level
sensor cord is properly inserted
into the connector on the sensor
and the connector on the back of
the OAS pump.
3. Verify that the red low saline LED
on the OAS pump control panel
turns off and either the yellow LED
or the green LED illuminates.
11 The OAS pump will
not power on
Note: The OAS pump is designed to
stop pumping saline and cut the
supply of power to the OAD if the OAS
pump encounters internal errors, such
as: improper OAS pump pumping
speed, problems with the power
supply, inadvertent opening of the
saline tubing door, etc. If an internal
error occurs, the green LED will turn
off and the yellow LED will illuminate.
1. Ensure that the saline tubing door
is closed.
2. Attempt to restart the OAS pump