Specifications

Diamondback360®CoronaryOrbitalAtherectomySystem
40
Issue
number
Issue Solution
saline tubing (i.e. bag spike) are
properly connected and a
sufficient amount of saline is in the
saline bag such that the low saline
level sensor is not active and the
red LED on the OAS pump control
panel is not illuminated.
3. Ensure that the OAS pump head
is properly aligned with the OAS
pump base.
4. Ensure that the saline tubing is
routed correctly through the OAS
pump saline tubing guides and
that the OAS pump saline tubing
door is closed.
5. Ensure that the yellow LED is off
and the green LED is illuminated.
If the green LED is not illuminated,
press the green Start button and
verify that the yellow LED is off
and that the green LED
illuminates.
8 The green (start)
LED is illuminated,
but the OAS pump
does not pump and
the yellow LED is
illuminated as well
Contact CSI Customer Service at the
phone number on the back of this
instructions for use.
9 The OAS pump
was running, but
has stopped
pumping and the
yellow LED is
illuminated
Press the Master Power switch, on
the back of the OAS pump, to power
off the OAS pump. Wait a few
seconds and press the Master Power
switch to power on the OAS pump.