Specifications
Diamondback360®CoronaryOrbitalAtherectomySystem
33
1. Open a fresh, sterile wipe that is pre-saturated with 70% Isopropyl
Alcohol (IPA) or prepare a sterile gauze/wipe by pouring or soaking it
with 70% IPA. Wring any excess IPA from the gauze/wipe, ensuring
that the gauze/wipe remains saturated, but not dripping, with IPA.
2. Thoroughly wipe all surfaces on the front face of the OAS pump.
Concentrate wiping the seams and crevices of the OAS pump head
door, around the edges of the lettering, and around the OAS pump
control panel. Continue to wipe these surfaces for a minimum of one
(1) minute. Discard the gauze/wipe. Repeat step #1.
3. Open the OAS pump head door. Using the prepared gauze/wipe,
thoroughly wipe the edge of the OAS pump head door on both sides
of the closure seam. Continue to wipe these surfaces for a minimum
of one (1) minute. Discard the gauze/wipe. Repeat step #1.
4. Thoroughly wipe all surfaces on the front face of the OAS pump and
the edge of the OAS pump head door on both sides of the closure
seam. Continue to wipe these surfaces for a minimum of one (1)
minute. Discard the gauze/wipe. Repeat step #1.
5. Continue to repeat this wiping process, as many times as necessary,
to ensure that all surfaces remain wet with IPA for a minimum of ten
(10) minutes.
6. Close the OAS pump door when disinfecting is complete.
10.4. Returning OAS components
Contact CSI Customer Service for returning OAS components. See the back of
this instruction for use for CSI contact information.