10.3.1. Cleaning the OAS Pump
Clean the OAS pump immediately after each use by following the steps
below:
Caution: Ensure that the OAS pump is powered off at the Master Power
switch on the back of the OAS pump and disconnect the OAS pump from
wall power before cleaning the OAS pump.
Caution: Do not immerse the OAS pump into fluids. Do not use solvents
or abrasive cleaners to clean the OAS pump as these may damage the
OAS pump and OAS pump components.
Caution: Completely dry the OAS pump before reconnecting the OAS
pump to wall power and powering on the OAS pump.
1. Prepare an enzymatic detergent, such as Enzol
®
, per manufacturer’s
directions.
2. Thoroughly wipe down the OAS pump, using a clean soft cloth that
has been dampened with the prepared detergent, until all visible soil
is removed.
3. Thoroughly rinse the OAS pump, using a clean soft cloth that has
been dampened with lukewarm tap water.
4. Dry the OAS pump using a clean, soft cloth; and, if available, filtered,
pressurized air at ≤40 psi.
10.3.2. Disinfecting the OAS Pump
Disinfect the OAS pump after each use by following the steps below:
Caution: Ensure that the OAS pump is powered off at the Master Power
switch on the back of the OAS pump and disconnect the OAS pump from
wall power before disinfecting the OAS pump.
Caution: Do not immerse the OAS pump into fluids. Do not use solvents
or abrasive disinfectants to disinfect the OAS pump as these may damage
the OAS pump and OAS pump components.
Caution: Completely dry the OAS pump before reconnecting the OAS
pump to wall power and powering on the OAS pump.